DC48 Automated sample
preparation

The DC48 supports the automation of liquid handling. With built-in barcode scanning, automatic addition of Proteinase K or internal standard, and setup of PCR reagents, the DC48 can be tailored to your specific needs. The 10.1’’ touch screen allows easy setup of programs and monitoring of the process in real time. The high-precision robotic arm controls the uncapping and recapping of your sample tubes. The DC48 includes several features to prevent cross-contamination, including only having one open sample tube at a time, a mobile drip tray following the robotic arm, a replaceable HEPA filter, and UV-disinfection lamps. The HEPA filters and UV disinfection are also key features in improving the biological safety of the operator.

Product Details

Reference Number:                                                         685

Product name                                                                  DC48                                                                                            Automated sample preparation

Intended Use

The DC48 supports the automaton of liquid handling. With built-in barcode scanning, automatic
addition of Proteinase K or internal standard, and setup of PCR reagents, the DC48 can be tailored to
your specific needs.
The 10.1’’ touch screen allows easy setup of programs and monitoring of the process in real time.
The high-precision robotic arm controls the uncapping and recapping of your sample tubes. The
DC48 includes several features to prevent cross-contamination, including only having one open
sample tube at a time, a mobile drip tray following the robotic arm, a replaceable HEPA filter, and
UV-disinfection lamps. The HEPA filters and UV disinfection are also key features in improving the
biological safety of the operator.

Product Specifications

Capacity                                                                           48 sample tubes

Pipetting volume                                                             5 – 1000 μL

Functions                                                                         1.Barcode scanning
                                                                                            2.Cap unscrewing
                                                                                            3.Cap rescrewing
                                                                                            4.Sample transfer
                                                                                            5.Reagent transfer
                                                                                            6.Liquid level detection
                                                                                            7.PCR-setup

Pipetting Performance                                                   Accuracy:
                                                                                            5-15 μL, CV=3%
                                                                                            15-50 μL, CV=1.5%
                                                                                            >50 μL, CV=1%
                                                                                            Repeatability:
                                                                                            5 ul~15ul: Relative deviation: ≤5%
                                                                                            15ul~50ul: Relative deviation: ≤3%
                                                                                            Above 50ul: Relative deviation: ≤1.5%

Run time                                                                           ~20 min

Sample Tubes                                                                  Tubes of 1mL, 3mL, 5mL and 10 mL

Interface option                                                               10.1 touch screen, Windows 10 in English, monitor all
                                                                                            the processes of pipetting and dispensing

Consumables                                                                   Prefilled 96 deepwell-plates(Nucleic acid extraction
                                                                                            kits) Stirring sleeves, black/conductive PCR single
                                                                                            tubes/strips/plates

Operating Power and Noise                                           AC 100-240V
                                                                                            50-60 Hz

Instrument Dimension and Weight                              67 cm (L) x 54 cm (W) x 74.2 cm (H)
                                                                                            80 kg (net)

Instructions For Use

The latest version of the Instructions For Use can be found at: 5ba.se/PP004 (ændre navnet (der er ikke noget link))